Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
Laboratory Procedure Manual
Analyte: HIV Antibody / HIV-1/HIV-2
Differentiation Assay
Matrix: Serum/Plasma
Method: Bio-Rad Multispot HIV-1/HIV-2 Rapid
Test
as performed by: HIV Laboratory Branch
Division of HIV/AIDS Prevention
National Center for HIV/AIDS, Viral Hepatitis, STD, and TB
Prevention
Contact: Tim Granade
Bill Switzer
First Published: October 2014
Revised:
Important Information for Users
CDC periodically refines these laboratory methods. It is the responsibility of the user to contact the
person listed on the title page of each protocol before using the analytical method to find out whether
any changes have been made and what revisions, if any, have been incorporated.
Public Release Data Set Information
This document details the Lab Protocol for testing the items listed in the following table:
File Name Variable Name SAS Label
HIV_I
LBXHIV1
LBXHIV2
HIV-1
HIV-2
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
1 of 24
Bio-Rad Multispot HIV-1/HIV-2 Rapid Test
1. Summary of Test Principle and Clinical Relevance- EIA
The Multispot HIV-1/HIV-2 Rapid Test is a single use qualitative immunoassay to detect and to
differentiate circulating antibodies to human immunodeficiency virus types 1 and 2 (HIV-1, HIV-2) in
fresh or frozen human serum and plasma. This rapid HIV-1/HIV-2 test kit is intended as an aid in the
diagnosis of infection with HIV-1 and/or HIV-2 in fresh or frozen human serum or plasma. This test is
suitable for use in multi-test algorithms designed for statistical validation of an HIV screening test
result or as part of an HIV-1/HIV-2 diagnostic testing algorithm that includes differentiation of HIV-1
and HIV-2 antibodies.
Summary and Explanation of the Test
Acquired immunodeficiency syndrome (AIDS) is caused by viruses transmitted by sexual contact,
exposure to blood (including sharing contaminated needles and syringes) or certain blood products, or
transmitted from an infected mother to her fetus or child during the perinatal period.
1
Additionally,
transmission of the viruses can occur through tissue transplantation.
2
HIV-1 has been isolated from
patients with AIDS and AIDS-related complex (ARC).
3-5
HIV-1 was thought to be the sole causative
agent of these syndromes until 1986, when a second type of HIV (HIV-2) was isolated and also
reported to cause AIDS.
6-7
Since the initial discovery, hundreds of cases of HIV-2 infection have been
documented worldwide, including cases of AIDS related to HIV-2.
8
In the United States, there have
been more than 80 cases of infection with HIV-2 reported, including three potential blood donors.
9-14
HIV-2 is similar to, but distinct from, HIV-1. Both viruses have similar morphology and
lymphotropism,
15
and the modes of transmission appear to be identical.
8,16
HIV-1 and HIV-2
genomes exhibit about 60% homology in conserved genes such as gag and pol, and 39-45%
homology in the envelope genes.
17
Serologic studies have also shown that the core proteins of HIV-
1 and HIV-2 display frequent cross-reactivity whereas the envelope proteins are more type-
specific.
18
Within the two major HIV types, there is significant variation as well. By analyzing sequences of
representative strains, HIV-1 has been divided into four groups: group M (for major), including at least
9 subtypes, 3 sub-subtypes of A, and 2 sub-subtypes of F (A1, A2, A3, B, C, D, F1, F2, G, H, J, and
K); group O (for outlier); group N (for non-M, non-O), and group P.
19-23
Similarly, HIV-2 strains have
been classified into at least five subtypes (A through E).
24
Some HIV-1 variants share ≤50% homology
in their env genes with those env sequences of more common prototype strains.
Despite some degree of immunological cross-reactivity between types and subtypes of HIV, reliable
detection of antibodies derived from the more divergent strains may only be achieved by incorporating
type- specific protein sequences into the assay design. In one study, detection of HIV-2-positive
samples by HIV-1 antibody kits ranged from 60% to 91%, depending on the test used.
25
The Multispot
HIV-1/HIV-2 Rapid Test incorporates highly conserved recombinant and synthetic peptide sequences
representing HIV-1 and HIV-2 envelope proteins.
26-32
The Multispot HIV-1/HIV-2 Rapid Test is
designed to detect antibodies to HIV-1 and HIV-2 in serum or plasma rapidly and reliably without
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
2 of 24
instrumentation. This test is suitable for use in multi-test algorithms designed for statistical validation
of rapid HIV test results or as part of an HIV-1/HIV-2 diagnostic testing algorithm that includes
differentiation of HIV-1 and HIV-2 antibodies
33-39
.
Biological Principles of the Procedure
The Multispot HIV-1/HIV-2 Rapid Test is based on the principle of ImmunoConcentration™.
40
The
Multispot HIV-1/HIV-2 Cartridge contains a removable specimen prefilter, the reaction membrane,
and an absorbent pad. All of the liquids added to the Cartridge are absorbed by the pad and contained
within the Cartridge. When the test is completed, the entire Cartridge can be decontaminated by
standard laboratory practices (see Precautions for Users) and properly discarded.
Microscopic particles are separately coated with the antigens that represent portions of the
transmembrane envelope proteins HIV-1 and HIV-2, respectively. The microparticles are
immobilized on the reaction membrane of the Multispot HIV-1/HIV-2 Cartridge and form the Test
Spots. The reaction membrane also contains a Procedural Control Spot that serves as a control spot
to ensure that the entire test procedure was properly executed. Samples to be tested are diluted in
Specimen Diluent and then added to the prefilter in the Cartridge. After the diluted specimen has
been completely absorbed, the prefilter is removed. If antibodies against HIV-1 and/or HIV-2 are
present in the specimen, they bind to the antigens on the microparticles in the specific spots on the
cartridge membrane. The Conjugate, which contains alkaline phosphatase-labeled goat anti-human
IgG (H+L chain specific), is then added to the Cartridge. The Conjugate binds to the human
antibody-antigen complexes that are immobilized in the spots on the cartridge membrane. Unbound
Conjugate is removed by a wash step.
Next, Development Reagent is added to the Cartridge. A purple color develops on the Test Spots in
proportion to the amount of antibodies against HIV-1 and/or HIV-2 that have been bound to the
antigen-coated microparticles and detected by the Conjugate. A purple color will also develop on
the Procedural Control Spot when the test has been performed correctly. Color development is
stopped by the addition of Stop Solution. The membrane is examined visually for the presence of
purple color on the Procedural Control Spot and on the Test Spots.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
3 of 24
2. Safety Precautions
1. This test kit should be handled only by qualified personnel trained in laboratory procedures and
familiar with their potential hazards. Handle appropriately with the requisite Good Laboratory
Practices including lab coat, eye/face protection and disposable gloves (synthetic, non-latex
gloves are recommended) while handling kit reagents and patient samples. Wash hands
thoroughly after performing the test.
2. Do not smoke, drink, or eat in areas where specimens or kit reagents are being handled.
3. Do not pipette by mouth.
4. This product contains dry natural rubber in the dropper bulb used with the Specimen Diluent
bottle.
5. Dispose of all specimens and materials used to perform the test as biohazardous waste.
Laboratory chemical and biohazardous wastes must be handled and discarded in accordance
with all local, regional, and national regulations. For additional information on biosafety
requirements, refer to CDC recommendations for Universal Precautions for prevention of
transmission of human immunodeficiency virus, hepatitis B virus, and other bloodborne
pathogens.
41
6. Complete hazard information and precautions are located in the Safety Data Sheet (SDS)
available at bio-rad.com or upon request.
Handling precautions:
1. Do not use any kit components beyond their stated expiration date.
2. Do not mix components from different lots.
3. Do not use the components in any other type of test kit as a substitute for the components in
this test kit.
4. Use the Multispot HIV-1/HIV-2 Cartridge and disposable Transfer Pipets only once and then
dispose of as described in Safety Precautions. Do not reuse these kit components.
5. Exercise care in opening and reusing reagent bottles to avoid microbial contamination of the
reagents.
6. Prior to running the assay, verify that the prefilter is seated firmly on top of the Cartridge
by pressing down firmly and evenly.
7. Always hold each reagent bottle vertically and allow each reagent to fall freely from the
dropper tip. Do not touch the dropper tip to any surface; this may contaminate the reagent.
8. Avoid contact of the Stop Solution with any oxidizing agent. Do not allow Stop Solution to
come into contact with metals.
9. Handle the Negative and Positive Control Serums in the same manner as patient specimens.
10. Inadequate adherence to package insert instructions may result in erroneous results.
11. When removing the Transfer Pipets from the bag, avoid touching the tips of the pipets.
12. The test should be performed with Cartridges that are placed on a flat surface.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
4 of 24
13. Adequate lighting is required to read test results.
WARNINGS FOR USERS
For In Vitro Diagnostic Use
1. The package insert must be read completely before performing the test. Failure to follow
the insert may give inaccurate test results.
2. The Bio-Rad Multispot HIV-1/HIV-2 Rapid Test has been approved for use with serum and
plasma specimens only. Use of this test kit with specimens other than those specifically
approved for use with this test kit may result in inaccurate test results.
3. Bring all reagents to room temperature (20-30°C) before use.
4. The following is a list of potential chemical hazards contained in some kit components
(refer to product REAGENTS chart):
a. WARNING: Some reagents contain 0.1% ProClin 150 or 0.5% ProClin 300:
H317: May cause an allergic skin reaction.
P280: Wear protective gloves/protective clothing/eye protection/face protection.
P302 + P352: IF ON SKIN: Wash with plenty of soap and water.
P333 + P313: If skin irritation or rash occurs: Get medical advice/ attention.
P501: Dispose of contents and container in accordance to local, regional, national and
international regulations.
ProClin 300 (0.1% ProClin 150 and 0.5% ProClin 300) are biocidal preservatives that are
irritating to eyes and skin, may be detrimental if enough is ingested, and may cause
sensitization by skin contact; prolonged or repeated exposure may cause allergic reaction in
certain sensitive individuals.
b. WARNING: Some reagents contain 0.1% Sodium Azide [NaN3]:
H303: May be harmful if swallowed.
H313: May be harmful in contact with skin.
P312: Call a POISON CENTER or doctor/physician if you feel unwell.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
5 of 24
Sodium azide may react with lead and copper plumbing to form metal azides that are highly
explosive. If disposed of in the sink, flush plumbing with a large volume of water to prevent
azide buildup.
c. The dilute 0.1 N sulfuric acid (H
2
SO
4
) Stop Solution may be detrimental if swallowed and by
contact, particularly to eyes. In case of contact with eyes, rinse immediately with plenty of
water and seek medical advice. Wastes can typically be neutralized to pH 6-8 for disposal if
trained and equipped to do so, however always dispose of dilute acidic / corrosive solutions in
accordance with local, regional, national and international regulations. Do not pour water into
this product.
5. Users of this test should follow the CDC Universal Precautions for prevention of transmission of
HIV, hepatitis B virus, and other bloodborne pathogens.
41
6. The Multispot HIV-1/HIV-2 Rapid Test contains human blood components. No known test
method can offer complete assurance that infectious agents are absent. Therefore, all human blood
derivates, reagents and human specimens should be handled as if capable of transmitting infectious
disease, following recommended Universal Precautions for bloodborne pathogens as defined by
OSHA, Biosafety Level 2 guidelines from the current CDC/NIH Biosafety in Microbiological and
Biomedical Laboratories
42
, WHO Laboratory Biosafety Manual
43
, and/or local, regional and
national regulations. The following human blood derivatives are found in this kit:
a. The Positive Control Serum has been heat-treated to inactivate HIV viruses and has been
tested and found nonreactive for Hepatitis B Surface Antigen (HBsAg) and antibodies (Ab) to
Hepatitis C virus (HCV).
b. The human source material used in the preparation of the Negative Control Serum has been
tested and found nonreactive for Hepatitis B Surface Antigen (HBsAg), antibodies to Hepatitis
C virus (HCV Ab), and antibodies to HIV (HIV-1/HIV-2 Ab).
Biological spills: Human source material spills should be treated as potentially
infectious.
Spills not containing acid should be immediately decontaminated, including the spill area,
materials and any contaminated surfaces or equipment with an appropriate chemical disinfectant
that is effective for the potential biohazards relative to the samples involved (commonly a 1:10
dilution of bleach, 70- 80% ethanol or isopropanol, an iodophor (such as 0.5% Wescodyne
TM
Plus), or a phenolic, etc.) and wiped dry.
Spills containing acid should be appropriately absorbed (wiped up) or neutralized, wiped dry and
then the area wiped with one of the chemical disinfectants; material used to absorb the spill may
require biohazardous waste disposal.
NOTE: DO NOT PLACE SOLUTIONS CONTAINING BLEACH INTO THE
AUTOCLAVE.
3. Computerization; Data System Management
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
6 of 24
HIV antibody differentiation results are manually entered into a Microsoft Excel result file
spreadsheet. After a run is complete and any additional corrections by the analyst are made, the
Excel result file is finalized. Data is transmitted electronically to Westat’s ISIS computer
system weekly and transferred from there to NCHS.
4. Specimen Collection, Storage, and Handling Procedures; Criteria for Specimen Rejection
Fresh or frozen serum or plasma collected by standard phlebotomy procedures may be used in
the test. The minimally acceptable volume of specimen available for performing the test is 40
µL. Approximately 30 µL is used for running each test. No clinically significant effect has
been detected in assay results of serum or plasma samples with increased levels of
hemoglobin, protein, albumin, lipids, or bilirubin. Performance of this assay has not been
evaluated on patient samples that have been heat-inactivated.
The following anticoagulants have been evaluated and found to be acceptable for use with
this test: EDTA, sodium citrate, sodium heparin, and SST tubes. Samples that are collected
into anticoagulant tubes should be filled as labeling indicates to avoid improper dilution.
Use of other anticoagulants has not been evaluated and may give incorrect results.
Specimens may be stored at 2-8°C for 7 days or at room temperature (20-30°C) for up to 48
hours. For long-term storage, the specimens should be frozen (-20°C or colder). Specimen
may be frozen and thawed up to 5 times.
If specimens are to be shipped, they should be packed in compliance with local regulations
covering the transportation of etiologic agents.
5. Procedures for Microscopic Examinations; Criteria for Rejection of Inadequately
Prepared Slides
Not applicable for this procedure
6. Equipment and Instrumentation, Materials, Reagent Preparation, Calibrators
(Standards), and Controls
A. Reagents
MULTISPOT HIV-1/HIV-2 Rapid Test
Product No. 25228 (50 Tests)
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
7 of 24
B. Reagent Preparation
All solutions and reagents are ready to use as supplied. Store kit at 2-8
o
C or room temperature (20-
30
o
C). If stored at 2-8
o
C, bring all reagents to room temperature before use, and return entire kit to
2-8
o
C when not in use. The kit may be used up to kit expiration date when stored at 2-8
o
C or for
up to 3 months if stored at room temperature. When stored at room temperature, change the
expiration date to three months after start of room temperature storage (do not change the date if it
less than 3 months before kit). Do not freeze test components.
C. Materials Provided
Package insert (1)
Subject Information Notice (1) - The Notice in the kit box may be copied as needed.
Customer letter (1)
D. Materials Required But Not Provided
1. Disposable glass or polypropylene test tubes (do not use polystyrene) to prepare
diluted specimens and controls (for example: 12 x 75 mm tubes)
Component
Contents
Preparation
1 • Multispot HIV-1/HIV-2
Cartridge
(50)
Foil-sealed base container with specimen
prefilter; Membrane with 1 Procedural Control
Spot and 3 Test Spots
Remove foil seal before use
2 • Positive Control
Serum
1 dropper bottle
(1 mL)
Heat-inactivated human serum/plasma
containing anti-HIV-1 and anti-HIV-2
immunoglobulin; Nonreactive for HBsAg and
antibody to HCV
0.1% Sodium azide
0.5% ProClin™ 300
Dilute in Specimen Diluent as
described
3 • Negative Control
Serum
1 dropper bottle
(1 mL)
Human serum; Nonreactive for HBsAg and
antibody to HIV and HCV
0.1% Sodium azide
0.5% ProClin™ 300
Dilute in Specimen Diluent as
described
4 • Specimen Diluent
1 dropper bottle
(25 mL)
Diluent for specimens and Controls
0.1% ProClin™ 150
0.125% ProClin™ 300
Dispense with dropper
provided
5 • Conjugate
1 dropper bottle
(9.5 mL)
Anti-human IgG (H+L) (goat) alkaline
phosphatase conjugated solution
0.1% ProClin™ 150
Ready to use as supplied
6 • Wash Solution
2 dropper bottles
(2 x 85 mL)
TRIS
Urea
Propylene glycol
Nitroblue tetrazolium
0.1% ProClin™ 150
Ready to use as supplied
7 • Development Reagent
1 dropper bottle
(8.5 mL)
3-Indoxyl phosphate
Ready to use as supplied
8 • Stop Solution
1 dropper bottle
(55 mL)
0.1 N H
2
SO
4
(sulfuric acid)
Ready to use as supplied
9 • Disposable Transfer
Pipets
(60)
Polyethylene transfer pipets
Ready to use as supplied
10 • Eyedropper (1)
Polyethylene eyedropper and cap with rubber
bulb; Contains Dry Natural Rubber, a potential
sensitizer
Use in Specimen Diluent
bottle
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
8 of 24
2. Test tube racks
3. Absorbent pads or paper towels
4. Biohazard bags with closures
5. Household bleach (5% or 8% sodium hypochlorite), diluted to a minimum
concentration of 10% bleach (0.5% sodium hypochlorite). Alternative disinfectants
include 70% ethanol or 0.5% Wescodyne™.
6. Disposable gloves.
7. Laboratory timer.
8. Precision pipettors that deliver 30 µL and 300 µL (optional for addition of specimen and
Specimen Diluent). Precision pipettors that deliver 10 µL and 90 µL as needed for dilutional
testing of dually positive samples.
9. Indelible laboratory marker.
7. Calibration and Calibration Verification Procedures
Procedural Control
Each Multispot HIV-1/HIV-2 Cartridge has a built-in procedural control, the Procedural
Control Spot, which is used to determine validity of the assay. The Procedural Control Spot
must be reactive (a definite purple spot) on each Cartridge for the results of that Cartridge to be
valid.
Quality Control
Using individual Multispot HIV-1/HIV-2 Cartridges as described in the Assay Procedure
above, run 1 Positive Control Serum and 1 Negative Control Serum (both provided in the kit)
under the following circumstances to monitor proper test performance:
A new operator uses the kit, prior to performing testing of specimens.
A new test kit lot is used.
A new shipment of kits is used.
The temperature used during storage of the kit falls outside of 2-30°C (35.6-86°F).
The temperature of the test area falls outside of 20-30°C (68-86°F).
According to intervals defined by the testing facility.
Results are read by examining the membrane and comparing the location of colored spots on the
membrane to the diagram below. Position the Multispot HIV-1/HIV-2 Cartridge with the ID label
facing the user. The appearance of any purple color in any of the Test Spots, regardless of intensity,
must be considered as presence of that Spot.
Expected results are as follows:
Negative Control
Note placement of Cartridge ID label
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
9 of 24
Only the Procedural Control Spot shows purple color development. The 3 Test Spots show no color
development.
. Positive Control
The Procedural Control Spot, both HIV-1 Test Spots, and the HIV-2 Test Spot show purple color
development.
Preliminary Statements
Once testing has been started, it should be completed without interruption.
Do not use more than ten (10) Multispot HIV-1/HIV-2 Cartridges in a batch, since using
more Cartridges may make it difficult to complete the testing without interruption. Larger
numbers of specimens can be tested by running several batches of up to 10 Cartridges.
The eyedropper used to dispense Specimen Diluent is packaged separately from the bottle
of Specimen Diluent. The first time a kit is used, remove the eyedropper from the
packaging and insert it into the bottle of Specimen Diluent. Discard the original cap and use
the eyedropper as the cap for the bottle. Two full eyedroppers dispenses approximately 300
μL of Specimen Diluent.
A 30μL precision pipettor can be used for addition of the sample to the Specimen Diluent.
The disposable Transfer Pipets supplied in the kit dispense approximately 30 μL per drop.
The Cartridges should be placed on a flat surface during the assay procedure to ensure
proper flow of specimen and reagents through the membrane.
All solutions must be completely absorbed (no standing liquid) into the Cartridge
membrane before proceeding to the next step in the Assay Procedure.
Assay Procedure
1. Bring kit and specimens to room temperature (20-30°C) before beginning
testing. It is essential that all kit components are at room temperature before use.
2. Place the required number of Cartridges on a flat surface with the patient ID label facing
toward the operator. Peel away the foil seals and discard them. Label the Cartridges to
correspond with the test tubes and the specimens to be tested.
Note: Verify that the blue prefilter and gray top support are seated securely in the base of
the Cartridge by pressing down firmly and evenly on both pieces. The prefilter must be
present in order to use the Cartridge for testing.
3. Label a test tube for each specimen or control to be tested.
4. Invert the Specimen Diluent bottle ten times to thoroughly mix just prior to drawing the
reagent.
5. Add two full eyedroppers of Specimen Diluent to each specimen and control tube.
Note: With the eyedropper in the Specimen Diluent, hold vertically and squeeze the bulb
completely, draw Specimen Diluent up into the eyedropper, and gently expel all of the
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
10 of 24
Specimen Diluent into the test tube. Repeat this sequence to deliver the second full
eyedropper.
6. Using a precision pipet with a separate pipet tip for each sample, add 30μL of specimen to
the Specimen Diluent. Alternatively, using a separate kit-provided Transfer Pipet for each
specimen, draw up a small amount of specimen. While holding the pipet vertically over the
appropriate dilution tube, add one drop to the tube.
Note: The drop should fall freely into the Specimen Diluent, not onto the side of the tube.
If the drop does fall onto the side of the tube, make sure that the entire drop drains down
into the Specimen Diluent. If the drop does not drain into the Specimen Diluent, discard
the tube and prepare a new dilution. Do not allow the tip of the pipet to touch any part of
the tube or the Specimen Diluent in the tube. Discard the used pipet tip or Transfer Pipet
into the biohazardous waste.
7. Test Positive and Negative Control Serums as described in the QC section. When
preparing Positive and Negative Control Serums, hold the dropper bottles vertically over
the tubes labeled for controls and squeeze gently.
Add one drop of each control to the appropriately labeled tube. The drop should fall freely
into the Specimen Diluent. Do not allow the tip of the dropper to touch any part of the
tube.
8. Mix each diluted specimen and control thoroughly. Mix gently to avoid foaming.
9. Pour the contents of each tube into the specimen prefilter of each corresponding pre-
labeled Cartridge, using a separate Cartridge for each tube. Wait two minutes, after which
the solution must be completely absorbed through the prefilter into the Cartridge.
10. Remove and discard the prefilter into the biohazardous waste.
11. Fill the central well of each Cartridge with Wash Solution by holding the bottle vertically
and squeezing gently. Do not touch bottle to solution in Cartridge well. Wait for the Wash
Solution to be absorbed completely before proceeding.
12. Add three drops of Conjugate to the central well of each Cartridge by holding the bottle
vertically and squeezing gently. Do not touch bottle tip to solution in Cartridge well. Wait
two minutes.
13. Fill the central well of each Cartridge with Wash Solution by holding the bottle vertically
and squeezing gently. Do not touch bottle to solution in Cartridge well. Wait for the Wash
Solution to be fully absorbed before proceeding.
14. Repeat previous step so that each Cartridge is washed twice. Wait for the Wash Solution to
be absorbed completely before proceeding.
15. Add three drops of Development Reagent to the central well of each Cartridge by holding
the bottle vertically and squeezing gently. Wait five minutes.
16. Fill the central well of each Cartridge with Stop Solution by holding the bottle vertically
and squeezing gently. Wait for the Stop Solution to be absorbed completely before reading
results.
17. Read test results according to Test Result Appearance and Interpretation, (Rapid Testing)
or (Antibody Differentiation Test in a Diagnostic Testing Algorithm), either immediately
or anytime up to 4 hours after completing the test. An elevated background can appear
over time with some specimens; therefore, reading results within 1 hour is optimal.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
11 of 24
Figure 1. Multispot Test procedure
1. Remove foil; press prefilter down. Label
cartridge and specimen or control tubes.
2. Add 2 full droppers of Specimen Diluent to
each microtiter tube.
3. Add one drop of each sample or control to each
labeled tube using a transfer pipette. Mix well.
Avoid foaming.
4. Pour each sample into the prefilter of the
labeled cartridge. Wait 2 minutes.
5. Remove and discard prefilter.
6. Fill the central well of each cartridge with
Wash Solution.
7. Once absorbed, add 3 drops of Conjugate. Wait
2 minutes.
8. Fill well with Wash Solution and let absorb.
Repeat.
9. Add 3 drops of Development Reagent.
Wait 5 minutes.
10. Fill well with Stop Solution. Allow to absorb
and read results.
Dilutional Procedure for HIV Differentiation Rapid HIV-1/HIV-2 Testing
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
12 of 24
The following procedure is used to differentiate samples that demonstrate purple color
development in the HIV-2 Spot as well as in one or both of the HIV-1 Spots.
1. Dilute the specimen 1:10 (using a calibrated pipettor, add 90 μL of Negative Control Serum
and 10 μL of sample to a separate test tube; or, alternatively, 135 μL of Negative Control
Serum and 15 μL of sample). Mix well.
2. Test the diluted sample as in steps 3-16 in the Assay Procedure section. Use the 1:10
diluted sample in place of the undiluted sample in step 6 of the Assay Procedure.
3. Read the results according to the criteria in Test Result Appearance and Interpretation.
If the results are nonreactive at this dilution, the specimen should be interpreted as
“Preliminary Positive for antibodies to HIV (undifferentiated).”
If the Procedural Control Spot is reactive and the Test Spots are reactive for only HIV-1 or
HIV-2, the result can be reported as Preliminary Positive for antibodies to the specific HIV
type identified.
If one or both of the HIV-1 Spots and the HIV-2 Spot are still reactive, continue testing as
follows.
4. Dilute the 1:10 diluted specimen again by 10-fold in Negative Control Serum, following the
procedure in step 1 above (the final dilution is 1:100).
5. Test the diluted sample as in steps 3-16 in the Assay Procedure section. Use the 1:100
diluted sample in place of the undiluted sample in step 6 of the Assay Procedure.
6. Read the results according to the criteria above in Test Result Appearance and
Interpretation.
If the Procedural Control Spot is reactive and the Test Spots are reactive for only HIV-1 or
HIV-2, the result can be reported as Preliminary Positive for antibodies to the specific HIV
type identified.
If the dual HIV reactivity does not disappear at the 1:100 dilution, or if the HIV-1 and HIV-
2 spots both become nonreactive at the same dilution, the specimen should be interpreted as
“Preliminary Positive for antibodies to HIV (undifferentiated).”
RAPID HIV-1/HIV-2 TESTING
Nonreactive:
Only the Procedural Control Spot shows purple color development. The 3 Test
Spots show no color development. Test result is interpreted as negative for HIV-1
and HIV-2 antibodies.
Reactive:
HIV-1 Reactive – Preliminary Positive:
The Procedural Control Spot shows purple color development and the recombinant
HIV-1 Spot and/or the HIV-1 Peptide Spot show purple color development. Test
result is interpreted as Preliminary Positive for HIV-1 antibodies.
HIV-2 Reactive – Preliminary Positive:
The Procedural Control Spot shows purple color development. The HIV-2 Peptide
Spot shows purple color development. Test result is interpreted as Preliminary
Positive for HIV-2 antibodies.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
13 of 24
HIV Reactive (Undifferentiated) Preliminary Positive:
The Procedural Control Spot shows purple color development. The HIV-2 Peptide
Spot shows purple color development as well as one or both HIV-1 Spots. In this
case, the specimen may be tested by additional methods which allow for
differentiation between HIV-1 and HIV-2. See dilutional procedure.
Invalid – Do Not Report Any Results
If no color develops in the Procedural Control Spot, regardless of color
development anywhere else on the membrane, the results are invalid.
If the background on the membrane is dark and interferes with the interpretation of
the spots, the results are invalid. In addition, if there are stray purple marks or
discoloration that interfere with reading the spots, the assay should be repeated.
Repeat the assay, and if results are still invalid collect a fresh sample or test by
another method.
If there is a halo of coloration that appears around all of the Test Spots and the
Procedural Control Spot, no results should be reported. An abnormal halo is
typically larger in size and lighter in color than the Procedural Control Spot. The
halo appears around each of the Test and Control Spots and may be apparent on the
membrane regardless of purple color development associated with positive results.
Samples exhibiting this unusual reaction should be tested by another methodology
to determine HIV status of the sample.
USE OF MULTISPOT AS AN ANTIBODY DIFFERENTIATION TEST IN A DIAGNOSTIC
TESTING ALGORITHM
Test Result Appearance and Interpretation Diagnostic Testing Algorithm that Includes
Differentiation between HIV-1 and HIV-2 Antibodies:
Place the Cartridges with the patient ID label facing toward the operator prior to reading test results.
Examine the Cartridge membrane and compare the location of colored spots to the diagram below.
The appearance of a definite purple color in any of the Test Spots must be considered as presence of
that Spot. Follow guidelines for using the assay in an HIV testing algorithm.
Nonreactive:
Only the Procedural Control Spot shows purple color development. The 3 Test
Spots show no color development. Test result is interpreted as negative for HIV-1
and HIV-2 antibodies. Additional testing is recommended, including HIV nucleic
acid testing (NAT).
Reactive:
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
14 of 24
HIV-1 POSITIVE:
The Procedural Control Spot shows purple color development and both the
recombinant HIV-1 Spot and the HIV-1 Peptide Spot show purple color
development. Test result is interpreted as Positive for HIV-1 antibodies
HIV-2 POSITIVE:
The Procedural Control Spot shows purple color development. The HIV-2 Peptide
Spot shows purple color development. Test result is interpreted as Positive for
HIV-2 antibodies.
HIV Reactive (Undifferentiated) :
The Procedural Control Spot shows purple color development. The HIV-2 Peptide
Spot shows purple color development as well as one or both HIV-1 Spots. In this
case, the specimen may be tested by additional methods which allow for
differentiation between HIV-1 and HIV-2. See dilutional procedure which
follows.
HIV-1 INDETERMINATE:
The Procedural Control Spot shows purple color development and either the
recombinant HIV-1 Spot or the HIV-1 Peptide Spot
shows purple color developmen
but not both HIV-1 Spots. Test result is interpreted as Indeterminate for HIV-1
antibodies and testing for HIV nucleic acid is recommended.
Invalid – Do Not Report Any Results
If no color develops in the Procedural Control Spot, regardless of color
development anywhere else on the membrane, the results are INVALID. (See
examples.)
If the background on the membrane is dark and interferes with the interpretation
of the spots, the results are invalid. In addition, if there are stray purple marks or
discoloration that interferes with reading the spots, the assay should be repeated.
Repeat the assay, and if results are still invalid collect a fresh sample or test by
another method.
If there is a halo of coloration that appears around all of the Test Spots and the
Procedural Control Spot, no results should be reported. An abnormal halo is
typically larger in size and lighter in color than the Procedural Control Spot. The
halo appears around each of the Test and Control Spots and may be apparent on
the membrane regardless of purple color development associated with positive
results. Samples exhibiting this unusual reaction should be tested by another
methodology to determine HIV status of the sample.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
15 of 24
Test Result Interpretation Dilutional Testing for Diagnostic Testing Algorithm
1:10 Dilutional Testing
Initial Result
1:10 Dilution Result
Interpretation/Action
HIV-1 POSITIVE
HIV-2 POSITIVE
Retest at 1:100 dilution
Retest at 1:100 dilution
HIV POSITIVE
(undifferentiated)
HIV-1 INDETERMINATE
HIV-2 POSITIVE
Retest at 1:100 dilution
HIV POSITIVE
(undifferentiated)
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
16 of 24
1:100 Dilutional Testing
Initial Result
1:10 Dilution Result
1:100 Dilution Result
Interpretation/Action
HIV-1 POSITIVE
HIV-2 POSITIVE
HIV POSITIVE
(undifferentiated)
HIV POSITIVE
(undifferentiated)
HIV POSITIVE
(undifferentiated)
HIV-1
INDETERMINATE
HIV-2 POSITIVE
HIV POSITIVE
(undifferentiated)
HIV POSITIVE
(undifferentiated)
8. Method Performance Specifications
Sensitivity for Antibodies to HIV-1
The sensitivity of the Multispot HIV-1/HIV-2 Rapid Test for antibodies to HIV-1 with serum
specimens is calculated to be 100% (95% CI = 99.94 – 100.00%).
The sensitivity of the Multispot HIV-1/HIV-2 Rapid Test for antibodies to HIV-1 with plasma
specimens is calculated to be 100% (95% CI = 99.94 – 100.00%).
Sensitivity for Antibodies to HIV-2
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
17 of 24
The sensitivity of the Multispot HIV-1/HIV-2 Rapid Test for antibodies to HIV-2 is calculated to be
100% (95% CI = 99.76 – 100%).
HIV-1 and HIV-2 Differentiation
Seven hundred ninety-nine (799) of the 801 samples (99.8%) were detected as HIV-1 reactive only on
Multispot HIV-1/HIV-2, and the remaining 2 samples were dually reactive (undifferentiated) on
Multispot HIV-1/HIV-2. Multispot identified 799 of the 801 known HIV-1 positive samples as HIV-1
reactive only (799/801 = 99.75%; 95% CI of 99.34 – 100.00%).
One hundred ninety (190) of 201 known HIV-2 specimens (94.5%) were detected as HIV-2 reactive
only on Multispot HIV-1/HIV-2. Nine were reactive for both HIV-1 and HIV-2 and two were
identified by Multispot as HIV-1 reactive.
Note: Samples reactive for both HIV-1 and HIV-2 may resolve to HIV-1 due to the lower titer of the
HIV-2 antibody as compared to the HIV-1 antibody. Dual infections with both HIV-1 and HIV-2
viruses are unusual but may occur in individuals from HIV-2 endemic countries.
Reactivity of Multispot HIV-1/HIV-2 on Worldwide Specimens and on HIV-1 Group O Serotype
Samples
A total of 79 frozen serum and 124 frozen plasma specimens from various worldwide geographic
locations outside of the U.S. were tested on Multispot HIV-1/HIV-2. HIV-1 subtypes represented
included subtypes A, B, C, D, E, F, and G. All 203 specimens from this worldwide panel were reactive
on Multispot HIV-1/HIV-2. In addition, 12 HIV-1 Serotype Group O frozen plasma samples were
tested on Multispot HIV-1/HIV-2. Ten (10) samples were from Cameroon, one was from Spain, and
one was from the United States. Eleven (11) of the 12 HIV-1 Group O serotype samples were reactive
when tested on Multispot HIV-1/HIV-2, and one was nonreactive.
Reactivity with Seroconversion and Sensitivity (Low and Mixed Titer) Panels
Multispot HIV-1/HIV-2 detected the presence of antibody to HIV-1 in specimens from ten
Seroconversion Panels as early as, or earlier than, a licensed HIV-1/HIV-2 EIA. The results of testing
Multispot HIV-1/HIV-2 on 2 low titer panels and 1 mixed titer panel, in comparison to a licensed HIV-
1/HIV-2 EIA, were assessed as well. Multispot HIV-1/HIV-2 was able to detect antibodies to HIV-1
similar to the licensed EIA.
Specificity
The specificity of the Multispot HIV-1/HIV-2 Rapid Test using serum specimens in these studies is
calculated to be 1494/1495 or 99.93% (95% CI = 99.79 – 100.00%).
The specificity of the Multispot HIV-1/HIV-2 Rapid Test using plasma specimens in these studies is
calculated to be 2272/2274 or 99.91% (95% CI = 99.77 – 100.00%).
Place the Cartridges with the patient ID label facing toward the operator prior to reading test results.
Examine the Cartridge membrane and compare the location of colored spots to the diagram below. The
appearance of any purple color must be considered as presence of that Spot.
9. Reportable Range of Results
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
18 of 24
Reportable results are expressed as positive or negative.
10. Quality Control (QC) Procedures
Procedural Control
Each Multispot HIV-1/HIV-2 Cartridge has a built-in procedural control, the Procedural
Control Spot, which is used to determine validity of the assay. The Procedural Control Spot
must be reactive (a definite purple spot) on each Cartridge for the results of that Cartridge to be
valid.
Quality Control
Using individual Multispot HIV-1/HIV-2 Cartridges as described in the Assay Procedure, run 1
Positive Control Serum and 1 Negative Control Serum (both provided in the kit) under the
following circumstances to monitor proper test performance:
A new operator uses the kit, prior to performing testing of specimens.
A new test kit lot is used.
A new shipment of kits is used.
The temperature used during storage of the kit falls outside of 2-30°C.
The temperature of the test area falls outside of 20-30°C.
According to intervals defined by the testing facility. The HIV Reference
Laboratory will include the Positive and Negative Control with each run.
Results are read by examining the membrane and comparing the location of colored spots on
the membrane to the diagram below. Position the Multispot HIV-1/HIV-2 Cartridge with the ID
label facing the user. The appearance of any purple color in any of the Test Spots, regardless of
intensity, must be considered as presence of that Spot.
Expected results are as follows:
Figure 2
Negative Control
Only the Procedural Control Spot shows purple color development. The 3 Test
Spots show no color development.
Positive Control
The Procedural Control Spot, both HIV-1 Test Spots, and the HIV-2 Test Spot
show purple color development.
11. Remedial Action If Calibration or QC Systems Fail To Meet Acceptable Criteria
a. Repeat the test if any of the controls do not meet expected reactivities.
b. Do not report any results from invalid runs
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
19 of 24
12. LIMITATIONS OF METHOD; INTERFERING SUBSTANCES AND CONDITIONS
1. For a preliminary positive result, when used as a rapid HIV-1/HIV-2 test, clinical
correlation is indicated with appropriate counseling, medical evaluation, and possibly
additional testing (for example, Western blot or indirect immunofluorescence assay) to
decide whether a diagnosis of HIV infection is accurate.
2. The Assay Procedure and the Test Result Appearance and Interpretation must be
followed closely when testing for the presence of antibodies to HIV-1 or HIV-2 in
plasma or serum from individual subjects. Failure to follow the procedure may give
inaccurate results.
3. The test was designed to test individual specimens of fresh or frozen serum or plasma.
Data regarding test kit interpretation were derived from testing individual samples.
Insufficient data are available to interpret tests performed on other body specimens,
pooled blood or processed plasma, and products made from such pools. Testing of these
specimens is not recommended.
4. The following anticoagulants have been evaluated and found to be acceptable for use with
this test: EDTA, sodium citrate, sodium heparin and SST tubes. Use of other
anticoagulants has not been evaluated and may give incorrect results.
5. Performance of this assay has not been evaluated on patient samples that have been
heat- inactivated.
6. Polystyrene tubes should not be used to prepare specimens for this test.
7. AIDS and AIDS-related conditions are clinical syndromes and their diagnosis can only be
established clinically. Testing alone cannot be used to diagnose AIDS, even if the
recommended investigation of reactive specimens suggests a high probability that the
antibody to HIV-1 or HIV-2 is present.
8. A nonreactive result for an individual subject indicates absence of detectable HIV
antibodies. However, a nonreactive test result does not preclude the possibility of
exposure to or infection with HIV-1 and/or HIV-2.
9. Nonreactive results can occur if the quantity of marker present in the sample is below
the detection limits of the assay, or if the marker that is detected is not present during
the stage of disease in which a sample is collected.
10. The risk of any asymptomatic person with a reactive serum or plasma developing AIDS
or an AIDS-related condition is not known, as the course of HIV infections may vary
among individual patients and may be altered by antiretroviral therapy. However, in a
prospective study, AIDS developed in 51% of homosexual men after 10 years of
infection.
43
11. A person who has antibodies to HIV-1 is presumed to be infected with the virus, except a
person who has participated in an HIV vaccine study may develop antibodies to the
vaccine and may or may not be infected with HIV.
12. Specimens which are reactive for antibodies to both HIV-1 and HIV-2 on initial testing should
be retested, according to the dilutional test protocol, to identify potential cross-reaction and
differentiate between HIV-1 and HIV-2. Results of dilutional testing, when used for rapid HIV-
1/HIV-2 testing, should be reported as Preliminary Positive for antibodies to the specific virus
type identified in the dilutional testing. Specimens that are dually reactive when tested
undiluted but only reactive for one virus type at the 1:100 dilution may be dually positive; these
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
20 of 24
samples are reported as Preliminary Positive for antibodies to the specific HIV type identified,
when used for rapid HIV-1/HIV-2 testing.
13. The intensity of the Test Spot does not correlate with antibody titer of the specimen.
14. Samples reactive for both HIV-1 and HIV-2 may resolve to HIV-1 at higher dilutions
due to the lower avidity of the HIV-2 antibody as compared to the HIV-1 antibody.
15. The Multispot HIV-1/HIV-2 Rapid Test cannot be used as part of a diagnostic testing
algorithm for both the initial testing and the differentiation testing of the same sample.
13. Reference Ranges (Normal Values)
A normal sample is negative for HIV antibodies.
14. Critical Call Results (Panic Values)
Not applicable to this assay method.
15. Specimen Storage and Handling during Testing
Specimens are stored at –20°C until testing. After an aliquot of the thawed sample has been
removed for testing, the residual is refrozen and stored at –20°C.
16. Alternate Methods for Performing Test or Storing Specimens if Test System Fails
If the analytical system fails, store specimens at –20°C until the system is investigated and the
problem is resolved. .
17. Test Result Reporting System; Protocol for Reporting Critical Calls (If Applicable)
Not applicable to this assay method.
18. Transfer or Referral of Specimens; Procedures for Specimen Accountability and Tracking
Standard record keeping involves using the computerized database and the hard copy results
themselves to track specimens. Records are maintained indefinitely. Only numerical identifiers
(e.g., case ID numbers) should be used. All personal identifiers should be available only to the
medical supervisor or project coordinator to safeguard confidentiality.
For the NHANES study, residual serum is retained at –70°C for 1 year and then returned to
NCHS serum bank.
19. Summary Statistics and QC Graphs
Qualitative assays are assays with a positive, negative or borderline/indeterminate result. Assay
controls are monitored for proper performance on each run.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
21 of 24
References
1. Centers for Disease Control: Provisional Public Health Service inter-agency recommendations
for screening donated blood and plasma for antibody to the virus causing acquired
immunodeficiency syndrome. Morbidity and Mortality Weekly Rep 34:5-7, 1985.
2. Delmonico FL, Snydman DR: Organ donor screening for infectious diseases. Transplantation
65(5):603-610, 1998.
3. Barre-Sinoussi F, Chermann JC, Rey F, et al: Isolation of a T-lymphotropic retrovirus
from a patient at risk for acquired immune deficiency syndrome (AIDS). Science 220:868-
871, 1983.
4. Gallo RC, Salahuddin SZ, Popovic M, et al: Frequent detection and isolation of cytopathic
retroviruses (HTLV-III) from patients with AIDS and at risk for AIDS. Science 224:500-503,
1984.
5. Coffin J, Haase A, Levy JA, et al: What to call the AIDS virus? Nature 321:10, 1986.
6. Clavel F, Guetard D, Brun-Vezinet F: Isolation of a new human retrovirus from West
African patients with AIDS. Science 233:343-346, 1986.
7. Clavel F, Manshino K, Chameret S, et al: Human immunodeficiency virus type 2
infection associated with AIDS in West Africa. New Engl J Med 316:1180-1185,
1987.
8. Schim van der Loeff MF and Aaby P: Towards a better understanding of the epidemiology of
HIV- 2. AIDS 13 (Suppl. A):S69-S84, 1999.
9. Centers for Disease Control: AIDS due to HIV-2 infection - New Jersey. Morbidity and
Mortality Weekly Rep 37:33-35, 1988.
10. Hoff R, Weiblen BJ, Schwerzler M, et al: Specific antibodies to HIV-2 detected in an
anonymous newborn blood specimen from Massachusetts. Fourth Consensus Conference on
Testing for Human Retroviruses, March 1989.
11. Ayanian JZ, Maguire JH, Marlink RG, et al: HIV-2 infection in the United States. New Engl J
Med 320:1422-1423, 1989.
12. O’Brien TR, George JR, Holmberg SD: Human immunodeficiency virus type 2 infection
in the United States. JAMA 267:2775-2779, 1992.
13. Centers for Disease Control: Update: HIV-2 infection among blood and plasma donors--
United States, June 1992-June 1995. Morbidity and Mortality Weekly Rep 44:603-606,
1995.
14. Sullivan PS, Fleming PL: Surveillance for HIV-2 in the United States: Update and
recommendations for future surveillance. Presented at the Association of Public
Health Laboratories Conference, Charlotte, NC, March 6-9, 2000.
15. Brun-Vezinet F, Katlama C, Roulot D, et al: Lymphadenopathy associated virus type 2 in
AIDS and AIDS-related complex. Lancet 1:128-132, 1987.
16. Quinn TC, Zacarias FRK, St. John RK: AIDS in the Americas: an emerging public health crisis.
New Engl J Med 320:1005-1007, 1989.
17. Guyader M, Emerman M, Sonigo P, et al: Genome organization and transactivation of the
human immunodeficiency virus type 2. Nature 326:662-669, 1987.
18. Cabrian K, Shriver K, Goldstein L, et al: Human immunodeficiency virus type 2: a
review. J Clinical Immunoassay 11:107-114, 1988.
19. Janssens W, Buvé A, Nkengasong JN: The puzzle of HIV-1 subtypes in Africa. AIDS 11:705-
712, 1997.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
22 of 24
20. Charneau P, Borman AM, Quillant C, et al: Isolation and envelope sequence of a highly
divergent HIV-1 isolate: definition of a new HIV-1 group. Virology 205:247-253, 1994.
21. Simon F, Mauclère P, Rogues P, et al: Identification of a new human immunodeficiency virus
type 1 distinct from group M and group O. Nature Medicine 4:1032-1037, 1998.
22. Meloni S T, et al. Distinct human immunodeficiency virus type 1 subtype A virus circulating
in West Africa: Sub-subtype A3. J Virology 78(22):12438-12445, 2004.
23. Plantier JC, Leoz M, Dickerson JE, De Oliveira F, Cordonnier F, Lemée V, Damond F,
Robertson DL, Simon F. A new human immunodeficiency virus derived from gorillas. Nat
Med. 8:871-2. 2009.
24. Gao, F, Yue L, Robertson DL, et al: Genetic diversity of human immunodeficiency virus
type 2: evidence for distinct subtypes with differences in virus biology. J Virology 68:7433-
7447, 1994.
25. George JR, Rayfield M, Philips S, et al: Efficacies of U.S. FDA-licensed HIV-1 screening
enzyme immunoassays for detecting antibodies to HIV-2. AIDS 4:321-326, 1990.
26. Starcich BR, Hahn BH, Shaw GM, et al: Identification and characterization of conserved and
variable regions in the envelope gene of HTLV-III/LAV, the retrovirus of AIDS. Cell 45:637-648,
1986.
27. Wang JJG, Steel S, Wisniewolski R, Wang CY: Detection of antibodies to human T-
lymphotrophic virus type III by using a synthetic peptide of 21 amino acid residues
corresponding to a highly antigenic segment of gp41 envelope protein. Proc Nat Acad Sci
USA 83:6159-6163, 1986.
28. Cosand WL: Synthetic antigen for the detection of AIDS-related disease. U.S. Patent
#4,629,783, 1986.
29. Fenouillet E, Sorensen A-M, Lacroix M, Coutellier A, Herson S, Fretz-Foucault C, Gluckman J-
C: Early and specific diagnosis of seropositivity to HIVs by an enzyme-linked immunosorbent
assay using env-derived synthetic peptides. AIDS 4:1137-41, 1990.
30. Gnann JW, McCormick, Mitchell S, Nelson J, Oldstone MBA: Synthetic peptide
immunoassay distinguishes HIV type 1 and HIV type 2 infections. Science 237:1346-1349,
1987.
31. Gnann JW, Nelson JA, Oldstone MBA: Fine mapping of an immunodominant domain in
the transmembrane glycoprotein of human immunodeficiency virus. J Virol 61:2639-
2641, 1987.
32. Alizon M, Wain-Hobson S, Montagnier L, Sonigo P: Genetic variability of the AIDS
virus: nucleotide sequence analysis of two isolates from African patients. Cell 46:63-
74, 1986.
33. Rosenberg, EC, Brennan CA, et al. Criteria for Laboratory Testing and Diagnosis of
Human Immunodeficiency Virus Infections; Approved Guideline Clinical and Laboratory
Standards Institute, M53-A. Vol 31. No 13. 2011.
34. Masciotra S, McDougal JS, Feldman J, Sprinkle P, Wesolowski L, Owen SM. Evaluation of an
alternative HIV diagnostic algorithm using specimens from seroconversion panels and persons
with established HIV infections. J Clin Virol. 2011 Dec;52 Suppl 1:S17-22.
35. Centers for Disease Control, Interpretation and Use of the Western Blot Assay for
Serodiagnosis of Human Immunodeficiency Virus Type 1 Infections. Morbidity and
Mortality Weekly Report 38(S-7): 1-7, 1989.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
23 of 24
36. Nasrullah M, Ethridge SF, Delaney KP, Wesolowski LG, Granade TC, Schwendemann J,
Boromisa RD, Heffelfinger JD, Owen SM, Branson BM. Comparison of alternative interpretive
criteria for the HIV-1 Western blot and results of the Multispot HIV-1/HIV-2 Rapid Test for
classifying HIV-1 and HIV-2 infections. J Clin Virol 2011 Dec.:S23-S27.
37. Styer LM, Sullivan TJ, Parker MM. Evaluation of an alternative supplemental testing strategy
for HIV diagnosis by retrospective analysis of clinical HIV testing data. J Clin Virol. 2011
Dec;52 Suppl 1:S35-40.
38. Torian LV, Forgione LA, Punsalang AE, Pirillo RE, Oleszko WR. Comparison of Multispot
EIA with Western blot for confirmatory serodiagnosis of HIV. J Clin Virol. 2011 Dec;52 Suppl
1:S41-44.
39. Wesolowski LG, Delaney KP, Hart C, Dawson C, Owen SM, Candal D, Meyer WA 3rd,
Ethridge SF, Branson BM. Performance of an alternative laboratory-based algorithm for
diagnosis of HIV infection utilizing a third generation immunoassay, a rapid HIV-1/HIV-2
differentiation test and a DNA or RNA-based nucleic acid amplification test in persons with
established HIV-1 infection and blood donors. J Clin Virol. 2011 Dec;52 Suppl 1:S45-9.
40. Valkirs GE, Barton R: ImmunoConcentration™ - A new format for solid phase immunoassays.
Clin Chem. 31: 1427-1431, 1985.
41. Centers for Disease Control: Universal precautions for prevention of transmission of human
immunodeficiency virus, hepatitis B virus, and other bloodborne pathogens in health-care settings.
Morbidity and Mortality Weekly Rep 37:377-388, 1988.
42. US Department of Health and Human Services. Biosafety in Microbiological and
Biomedical Laboratories. 5
th
ed. Washington, DC: US Government Printing
Office, HHS publication No. (CDC) 21-1112. December 2009.
43. World Health Organization. Laboratory Biosafety Manual. 3
rd
ed. Geneva: World
Health Organization. 2004.
Multispot HIV-1/HIV-2 Rapid Test
NHANES 2015-2016
24 of 24