restriction or loss of material. A splitter
brush should be used to clean the chutes of
adhering fines.
NOTE 1 - Mechanical splitters are
commonly available in sizes adequate for
coarse aggregate having the largest particle
not over 1 1/2 in. (37.5 mm).
11. PROCEDURE
11.1 The riffle splitter must be clean and
dry before use. Place the material into a
large, flat bottomed mixing pan. Mix the
material thoroughly. Using a flat scoop
equal in width to the overall length of the
riffles, remove material from the pan and
slowly pour the material into the riffle splitter
first from one side and then the other.
Alternatively, use a flat, square end scoop to
load the sample from the mixing pan into
two extra splitter pans placed side by side.
Slowly pour approximately half of the
sample in the pan from one side and then
reverse the ends of the pan and pour the
remainder from the other side. A slight
jarring action by the pan against the splitter
helps keep the riffles from clogging.
Uniformly distribute the sample from edge to
edge, so that when it is introduced into the
chutes, approximately equal amounts will
flow through each chute. The rate at which
the sample is introduced shall be such as to
allow a free flow through the chutes into the
receptacles below. Do not allow any of the
riffles to become plugged since this will
divert material to the two adjacent riffles and
send too much material to the opposite
receiving pan.
11.2 Reintroduce the portion of the
sample from alternating receptacles into the
splitter as many times as necessary to
reduce the sample to the size specified for
the intended test. Retain the portion of the
material collected in the other receptacle at
the last split until tests are completed.
NOTE 2 - As an alternative to Section 11.2,
further splitting to testing size can be
achieved with Section 11.3.
11.3 After splitting the material into two or
four equal measures (depending on the size
of the field sample), leave the divided
sample in the splitter pans and place in the
oven. Use the flat, square end scoop to
obtain individual test samples of the required
weight. Work from one end of the pan to the
other. Insert the scoop to the full depth of
the material. Every attempt should be made
to minimize the loss of particles over the
sides of the scoop. A putty knife may be
used to separate the material in the scoop
from the material in the pan and also to cut
increments from the main body of material in
the scoop. Do not shake the material in the
scoop to add small, additional amounts to
the specimen, as this may introduce
segregated material to the specimen. Save
the remaining portion of the sample until
tests are completed.
METHOD D -
SELECTION BY CROSS SECTION
12. APPARATUS
12.1 Apparatus shall consist of a small,
flat, square end scoop with square sides; a
putty knife; and two slats having a height at
least one inch taller than the sides of the
splitting pan. The slats shall conform within
one inch to the sides of the pan, so that
material can not fall from the vertical face
into the sample being separated.
13. PROCEDURE
13.1 Prepare the sample for reduction
per Section 5.1 or Section 5.2.
13.2 Obtain a sample for each test by
pushing a dividing slat vertically through the
entire width of the sample until it contacts
the bottom of the pan. Next, place a second
slat parallel to the first and push it vertically
to the bottom of the pan. Remove all of the
material between the slats. Take care to
include all fines from the pan, the slat sides,
and the utensil in the sample. Obtain
additional samples by pushing one of the
slats vertically into the remaining material
and repeating the process. Save the
remaining portion of the sample until tests
are completed.